gains and offsets using programmable DACs. On these devices,
\fBcomedi_calibrate\fR adjusts the programmable calibration
DACs so that the measured value of reference voltages fall
-within acceptable limits.
+within acceptable limits. The calibration settings are
+saved to a file which can be fed to the comedi_apply_calibration()
+function.
\fBComedi_calibrate\fR prints a lot of debugging information
as it runs.
\fBcomedi_calibrate\fR recognizes the following options:
+\fB--[no-]calibrate\fR Perform calibration (or not).
+
\fB-f, --file /dev/comediN\fR perform calibration on device \fB/dev/comediN\fR
-\fB--help print help for options then exit
+\fB--help\fR print help for options then exit
+
+\fB-q, --quiet\fR don't generate output to stdout
-\fB-q, --quiet don't generate output to stdout
+\fB--[no-]reset\fR Reset (or don't reset) calibration subdevice to
+a reasonable starting point before
+performing calibration. The default is \fB--reset\fR. The
+\fB--no-reset\fR option is useful if
+a single calibration
+pass is not sufficient, and you wish to perform another calibration
+pass starting from the results of the previous pass.
\fB-s, --save-file <file path>\fR save calibration information to specified file,
instead of default location