Boards may have one of 4 calibrations statuses, depending
on how well the calibration code is trusted. These are:
-STATUS_NONE, the default for no information; STATUS_SOME,
+STATUS_UNKNOWN, the default for no information; STATUS_SOME,
meaning that a dump has been converted to initial code,
but not tested; STATUS_DONE means that the output of a
STATUS_SOME dump has been checked, and is correct;
The unipolar zero offset may not be available on some boards.
-In a STATUS_NONE dump, for each measurable quantity and each
+In a STATUS_UNKNOWN dump, for each measurable quantity and each
calibration DAC, the DAC is varied throughout its entire range
and the quantity measured. The data is linearly fit, and if
the slope is statistically non-zero, a line is printed:
E series boards, so I'm not too worried about the following
dump where there are S_min/dof ratios above 4.
-Here's an example dump, generated by a STATUS_NONE dump for
+Here's an example dump, generated by a STATUS_UNKNOWN dump for
a pci-mio-16xe-10, with the analog output section removed:
Warning: device not fully calibrated due to insufficient information